Single Bag for Ascites Drainage

Single Bag for Ascites Drainage
Sterile single bag for collection of ascites fluid, packed in a single overwrap.

Composition

See brochure.

Capacity

Graduated bags, 200ml to 3500ml.

Theraputic indications

Bag for drainage of ascites.

Sterilisation Method.

Sterilisation by irradiation in the overwrap. NB: The bags can be slightly coloured without this being a visible sign of deterioration.

Storage Conditions and Shelf-life.

  • Store out of direct sunlight and frost at a temperature between 0░C and +35░C.
  • 5 years shelf life in sealed packaging.

Compliance

Sterile class I device (compliant with EC labelling dossier).

Instructions for use

Preparation

  • Before use, check the expiry date; ensure that the overwrap and the device are intact. Reject any faulty contents. If necessary, bring to a temperature between +15░C and +30░C.
  • Prepare the material necessary for the ascites puncture: bags, needles, skin antiseptic, pressure pads, catheterů
  • Only remove the device from its packaging just before use. Place the bags onto a clean surface (the overwrap may be used for this purpose). Close the blue clamps and the flow regulator.
  • 2.Puncture

  • Proceed with the puncture in an aseptic manner, according to musual procedure.
  • Connect the luer-lock to the device at the puncture site.
  • Open the clamp and the flow regulator. The ascites liquid flows into the tubing. If necessary, prime the device by connecting a syringe to the 3-way stopcock.
  • It is possible to proceed with the collection of ascites liquid in the same way.
  • Fill the drip chamber.
  • Place the bag onto a clean surface or hang it by its eyelets, 30cm below the puncture site.
  • Establish the quantity of ascites liquid collected by reading the graduated scale.
  • If the volume collected is above to 3500ml, an additional bag can be connected (ref VSL8000XQ). See the relevant technical data file for instructions on how to use.
  • 3.Disposal

  • At the end of the puncture, close the flow regulator and the clamps.
  • Remove the device and dispose of it in accordance with the regulations in place concerning high-risk waste.

Packaging

Product referenceN░ of devices per packageN░ of devices per box
VSL8003XQ125

Each bag is individually wrapped.